Is the correlation between fetal oxygen saturation and blood pH sufficient for the use of fetal pulse oximetry?

R. J P Rijnders, B. W J Mol, P. J H M Reuwer, A. P. Drogtrop, M. M A Vernooij, G. H A Visser

Research output: Contribution to journalArticle

14 Citations (Scopus)

Abstract

Objectives: Fetal pulse oximetry was performed during labor in high-risk cases for fetal distress to determine the diagnostic value of this method. Methods: The fetal SpO2 values were blinded from the obstetrician so that these values did not influence clinical decisions. Mean and lowest SpO2 measurements for the last 30 min prior to either fetal scalp blood sampling or delivery were correlated with scalp pH or pH from the umbilical artery. Results: No significant correlation was found between pH and mean fetal oxygen saturation (correlation coefficient -0.02, p = 0.9). There was no significant correlation between pH and lowest fetal oxygen saturation (correlation coefficient 0.04, p = 0.84). Concerning the feasibility of the method, we found that only 23 of 65 included patients were suitable for analysis; in 20% of cases, we were not able to perform a SpO2 measurement. Conclusions: None of three cases with pH below 7.05 would have been detected using mean SpO2 over the last 30 min prior to fetal scalp blood sampling or delivery. Only one case would have been detected using the lowest SpO2 measurement over this period. We conclude that fetal SpO2 measurements during labor are of poor diagnostic value, with a disappointing feasibility and therefore are not ready for implementing into daily clinical practice.

LanguageEnglish
Pages80-83
Number of pages4
JournalJournal of Maternal-Fetal and Neonatal Medicine
Volume11
Issue number2
Publication statusPublished - 2002

Keywords

  • Diagnostic value
  • Feasibility
  • Fetal monitoring
  • Fetal pulse oximetry

ASJC Scopus subject areas

  • Pediatrics, Perinatology, and Child Health
  • Obstetrics and Gynaecology

Cite this

Rijnders, R. J. P., Mol, B. W. J., Reuwer, P. J. H. M., Drogtrop, A. P., Vernooij, M. M. A., & Visser, G. H. A. (2002). Is the correlation between fetal oxygen saturation and blood pH sufficient for the use of fetal pulse oximetry? Journal of Maternal-Fetal and Neonatal Medicine, 11(2), 80-83.
Rijnders, R. J P ; Mol, B. W J ; Reuwer, P. J H M ; Drogtrop, A. P. ; Vernooij, M. M A ; Visser, G. H A. / Is the correlation between fetal oxygen saturation and blood pH sufficient for the use of fetal pulse oximetry?. In: Journal of Maternal-Fetal and Neonatal Medicine. 2002 ; Vol. 11, No. 2. pp. 80-83.
@article{a49a8ae4fcf7437fbe70c2854c20caa6,
title = "Is the correlation between fetal oxygen saturation and blood pH sufficient for the use of fetal pulse oximetry?",
abstract = "Objectives: Fetal pulse oximetry was performed during labor in high-risk cases for fetal distress to determine the diagnostic value of this method. Methods: The fetal SpO2 values were blinded from the obstetrician so that these values did not influence clinical decisions. Mean and lowest SpO2 measurements for the last 30 min prior to either fetal scalp blood sampling or delivery were correlated with scalp pH or pH from the umbilical artery. Results: No significant correlation was found between pH and mean fetal oxygen saturation (correlation coefficient -0.02, p = 0.9). There was no significant correlation between pH and lowest fetal oxygen saturation (correlation coefficient 0.04, p = 0.84). Concerning the feasibility of the method, we found that only 23 of 65 included patients were suitable for analysis; in 20{\%} of cases, we were not able to perform a SpO2 measurement. Conclusions: None of three cases with pH below 7.05 would have been detected using mean SpO2 over the last 30 min prior to fetal scalp blood sampling or delivery. Only one case would have been detected using the lowest SpO2 measurement over this period. We conclude that fetal SpO2 measurements during labor are of poor diagnostic value, with a disappointing feasibility and therefore are not ready for implementing into daily clinical practice.",
keywords = "Diagnostic value, Feasibility, Fetal monitoring, Fetal pulse oximetry",
author = "Rijnders, {R. J P} and Mol, {B. W J} and Reuwer, {P. J H M} and Drogtrop, {A. P.} and Vernooij, {M. M A} and Visser, {G. H A}",
year = "2002",
language = "English",
volume = "11",
pages = "80--83",
journal = "Journal of Maternal-Fetal and Neonatal Medicine",
issn = "1476-7058",
publisher = "Informa Healthcare",
number = "2",

}

Rijnders, RJP, Mol, BWJ, Reuwer, PJHM, Drogtrop, AP, Vernooij, MMA & Visser, GHA 2002, 'Is the correlation between fetal oxygen saturation and blood pH sufficient for the use of fetal pulse oximetry?', Journal of Maternal-Fetal and Neonatal Medicine, vol. 11, no. 2, pp. 80-83.

Is the correlation between fetal oxygen saturation and blood pH sufficient for the use of fetal pulse oximetry? / Rijnders, R. J P; Mol, B. W J; Reuwer, P. J H M; Drogtrop, A. P.; Vernooij, M. M A; Visser, G. H A.

In: Journal of Maternal-Fetal and Neonatal Medicine, Vol. 11, No. 2, 2002, p. 80-83.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Is the correlation between fetal oxygen saturation and blood pH sufficient for the use of fetal pulse oximetry?

AU - Rijnders, R. J P

AU - Mol, B. W J

AU - Reuwer, P. J H M

AU - Drogtrop, A. P.

AU - Vernooij, M. M A

AU - Visser, G. H A

PY - 2002

Y1 - 2002

N2 - Objectives: Fetal pulse oximetry was performed during labor in high-risk cases for fetal distress to determine the diagnostic value of this method. Methods: The fetal SpO2 values were blinded from the obstetrician so that these values did not influence clinical decisions. Mean and lowest SpO2 measurements for the last 30 min prior to either fetal scalp blood sampling or delivery were correlated with scalp pH or pH from the umbilical artery. Results: No significant correlation was found between pH and mean fetal oxygen saturation (correlation coefficient -0.02, p = 0.9). There was no significant correlation between pH and lowest fetal oxygen saturation (correlation coefficient 0.04, p = 0.84). Concerning the feasibility of the method, we found that only 23 of 65 included patients were suitable for analysis; in 20% of cases, we were not able to perform a SpO2 measurement. Conclusions: None of three cases with pH below 7.05 would have been detected using mean SpO2 over the last 30 min prior to fetal scalp blood sampling or delivery. Only one case would have been detected using the lowest SpO2 measurement over this period. We conclude that fetal SpO2 measurements during labor are of poor diagnostic value, with a disappointing feasibility and therefore are not ready for implementing into daily clinical practice.

AB - Objectives: Fetal pulse oximetry was performed during labor in high-risk cases for fetal distress to determine the diagnostic value of this method. Methods: The fetal SpO2 values were blinded from the obstetrician so that these values did not influence clinical decisions. Mean and lowest SpO2 measurements for the last 30 min prior to either fetal scalp blood sampling or delivery were correlated with scalp pH or pH from the umbilical artery. Results: No significant correlation was found between pH and mean fetal oxygen saturation (correlation coefficient -0.02, p = 0.9). There was no significant correlation between pH and lowest fetal oxygen saturation (correlation coefficient 0.04, p = 0.84). Concerning the feasibility of the method, we found that only 23 of 65 included patients were suitable for analysis; in 20% of cases, we were not able to perform a SpO2 measurement. Conclusions: None of three cases with pH below 7.05 would have been detected using mean SpO2 over the last 30 min prior to fetal scalp blood sampling or delivery. Only one case would have been detected using the lowest SpO2 measurement over this period. We conclude that fetal SpO2 measurements during labor are of poor diagnostic value, with a disappointing feasibility and therefore are not ready for implementing into daily clinical practice.

KW - Diagnostic value

KW - Feasibility

KW - Fetal monitoring

KW - Fetal pulse oximetry

UR - http://www.scopus.com/inward/record.url?scp=0035988739&partnerID=8YFLogxK

M3 - Article

VL - 11

SP - 80

EP - 83

JO - Journal of Maternal-Fetal and Neonatal Medicine

T2 - Journal of Maternal-Fetal and Neonatal Medicine

JF - Journal of Maternal-Fetal and Neonatal Medicine

SN - 1476-7058

IS - 2

ER -